Imaging Artifacts

mirror artifact

Ultrasound Assumptions

2D Imaging artifacts are causes by violation of several ultrasound assumptions. The table below describes this:

Ultrasound AssumptionArtifact Generated

Pulses and echoes travel in straight lines
Refraction

Waves are infinitely thin or pulses extremely small
Beam width

Echoes travel back to the transducer after a single reflection
Reverberation, mirror image

Echoes originate from the main beam
Side lobe, grating lobe

Speed of sound in tissues is constant
Refraction

Attenuation is uniform by all tissues
Acoustic shadowing, enhancement
Anesthesia & AnalgesiA 2016

Dealing With Artifacts

  • Shadowing refers to hypoechoic or anechoic areas distal to a STRONG reflector. (eg rib shadows)
  • Enhancement refers to hyperechoic areas distal to a WEAK reflector. (eg bladder or fluid filled cyst)
  • Reverberation – Multiple duplicated images spaced at regular intervals. (eg. A-Lines)
  • Refraction – Misregistration, ommision or duplication of an object.
  • Mirror image – Duplicated image deep and equidistant from object
  • Beam width – Lateral blurring of an image, 2 objects may appear as one.
  • Side lobe – Blurring of rdges of image.

Most artifacts can be dealt with by moving the transducer so that the beam does not pass the reflector causing the issue. Scanning in 2 planes should remove artifacts. Optimizing, depth, gain and focal zone are helpful.

References

Imaging artifacts in echocardiography. Le et al. Anaesthesia & Analgesia 2016.

Basic Ultrasound Physics Course