Introduction
This can cause confusion in image interpretation. Knowledge of artifacts and why they happen can help us to avoid pitfalls in diagnosis and improve imaging quality. We hope you find this summary of what is a difficult topic useful !
Classification of Artifacts
Artifacts can be calssified into:
- Artifacts related to ultrasound beam properties.
- Artifacts related to wave reflection or refraction.
- Doppler artifacts – these are not covered presently.
1. Artifacts Related to Ultrsound Beam Properties
a) Side Lobe Artifact
When a side lobe beam encounters a STRONG REFLECTOR, the reflected side lobe beam will be interpreted by the processing functions of the machine as a reflection from the main beam (where the transducer is ‘looking’). This results in the display of the reflector along the path of the main beam instead of its anatomical location.
b) Beam Width Artifact
This is caused by the violoation of the assumption that the ultrasound beam is infinitely thin. In reality, the lateral and elevational width of the beam give the ultrasound beam a 3-Dimensional shape. Anatomical structures outside of the 2D imaging plane are then incorrectly positioned as if they are within the scanning plane on the display monitor.
c) Near Field Clutter
Structures in the near field are sometimes obscured due to the high amplitude of oscillations by the transducer itself, causing a so called “near field clutter”.
2. Artifacts Related to Wave Reflection and Refraction
a) Mirror Image
A mirror image artifact typically appears below a strong reflective surface producing a duplicate image behind the mirror of the real structures in front of the mirror.
b) Acoustic Shadowing
Acoustic shadowing results in the absence of echoes behind a STRONG reflector. This is due to a strong reflector or refractor preventing ultrasound wave propagation beyond it.
c) Acoustic Enhancement
When the ultrasound beam encounts a weakly attenuating structure, the echoes received from points distal to this material are higher in intensity than echoes received from a similar depth in the imaging plane. This results in a brighter display of the image distal to the weak attenuator. This is characteristic of fluid-filled structures such as cysts and the gallbladder.
d) Edge Shadowing
e) Reverberation Artifact
These are ghost images caused by reflection of ultrasound waves between 2 strong reflectors. This causes delayed return of the beam to the transducer leading the machine to interpret the presence of multiple ghost images at equidistant intervals below the true image. A Lines seen in a well aerated lung are a classic example. These are caused by repeated reflection of the ultrasound beam between the pleural surface and the trasnducer head.
Reverberation artifacts can be reduced or avoided by reducing the gain or scanning at different angles. A lines seen in lung ultrasound signify well aerated lungs. They are reverberation artifacts that will not be seen if the angle between the ultrasound beam and pleural surface is not perpendicular.
f) Comet Tail
Fig 1.8. Comet tail artifacts are also seen in places such as the descending aorta where they can be seen to radiate from the distal wall. They are caused by closely spaced reflectors, in the case of the aorta – this is due to aortic wall calcification. Image from: Imaging Artifacts in Echocardiography. Le et al. A&A 2016.
g) Ring Down Artifact (B Lines)
Small pockets of fluid trapped by surrounding air bubbles can cause ring down artifact. When ultrasound waves hit a pocket of trapped fluid, the sound waves resonate within the fluid before being transmitted back to the transducer. The resonant vibrations detected by the ultrasound transducer are displayed as bright vertical lines extending deep to the trapped fluid.
Video: Ultrasound Artifacts
Further Reading
- Imaging Artifacts in Echocardiography. Le et al. Anesthesia & Analgesia 2016.
- Fact or Artifact in Two-Dimensional Echocardiography: Avoiding Misdiagnosis and Missed Diagnosis. Bertrand et al. JASE 2016.
- US Artifacts. Feldman et al. RSNA. 2009